Orthovoltage X-rays : ウィキペディア英語版
Orthovoltage X-rays

Orthovoltage x-rays are produced by x-ray generators operating at voltages in the 200–500 kV range, and therefore an energy in the 200–500 keV range (see external beam radiotherapy for an explanation of the maximum and mean energies as a function of voltage). When used to treat patients, radiation oncologists find that they penetrate to a useful depth of about 4–6 cm.〔Advances in kilovoltage x-ray beam dosimetry http://iopscience.iop.org/0031-9155/59/6/R183/article〕 That makes them good for treating skin, superficial tissues, and ribs, but not for deeper structures such as lungs or pelvic organs.
Orthovoltage X-rays are sometimes termed "deep" x-rays (DXR). The energy and penetrating ability of the x-rays produced by an x-ray tube increases with the voltage on the tube. External beam radiotherapy (teletherapy) began around the turn of the 20th century with ordinary diagnostic x-ray tubes, which used voltages below 150 kV. Physicians found that these were adequate for treating superficial tumors, but not tumors inside the body. Since these low energy x-rays were mostly absorbed in the first few centimeters of tissue, to deliver a large enough radiation dose to buried tumors would cause severe skin burns.
Therefore beginning in the 1920s "orthovoltage" 200–500 kV x-ray machines were built. These more penetrating x-rays were termed "deep" or "skin sparing" rays. These were found to be able to reach shallow tumors, but to treat tumors deep in the body more voltage was needed. By the 1930s and 40s huge megavoltage x-ray machines, using 3-5 million volts on the tube, began to be built. With the introduction of linear accelerators in the 1970s, which could produce 4-30 MV beams, orthovoltage x-rays are now considered quite shallow.
By convention, the voltage is used to characterize X- and gamma-ray beams (in volts), while electron beams are characterized by their energies (in electronvolts).
==See also==

* Megavoltage X-rays